25
frequency component which is higher than 20MHz obstructed.
Press“CH1”“BW Limit” “Off”, Set bandwidth Limited to close state, the High
Frequency component in the tested signal could pass.
Picture 2.6-3
3. Adjust Sensitivity
Vertical scale adjust have Coarse and fine two modes, Vertical sensitivity range of
SDS1000C/CE/CF/CFL are 2mV/div5V/div scale. Vertical sensitivity range of
SDS1000CM/CN/D is 2mV/div10V/div scale .
Take the CH1 for example:
Press“CH1”“Volts/Div”“Coarse”. It is the default setting of Volts/Div, and it makes
the vertical scaling in a 1-2-5-step sequence from 2mv/div, 5mv/div, 10mv/div to 5v/div.
Press CH1” Volts/Div” Fine”. This setting changes the vertical to small steps
between the coarse settings. It will be helpful when you need to adjust the waveform
vertical size in smooth steps.
Set bandwidth to
20MHZ
Bandwidth Limit
Symbol