2.13.6 Updating the System Software
Using USB flash drive updat frimware
The software of the oscilloscope can be updated directly via USB Flash drive, this
process needs about two minutes. Follow next steps:
1. Insert USB Flash Drive with firmware procedure to USB Host interface on the
front panel of the oscilloscope.
2. Press the “UTILITY” button to enter the “Utility menu”.
3. Press “Next Page” option button to enter the third page of “Utility menu”.
4. Press the “Update Firmware” option button.
5. Press the “SINGLE” button to start updating the software according to prompt
on the screen.
Turn off the oscilloscope and turn on it again, the software has been updated. The
oscilloscope
needs to execute “Do Self Cal” after update.
Note: Don’t cut off the power while oscilloscope is updating.
Ethernet port upgrate function
Users can upgrate the DSO through the Ethernet port on the DSO back pannel.
operation
Steps are as follows
1. open the DSO
2. connect the DSO to PC with the net line
3. set the IP address of the PC as10.11.0.110
4. Netmask 255.255.255.0
5. Gateway 10.11.0.1
6. pressUTILITYbutton to enter the third page,chooseIP Setting
7. adjust the universal knob to set the DSO IP as10.11.0.116
8. Net Mask255.255.255.0