Strany 1 - User Manual

I User Manual SDS1000 Series Digital Storage Oscilloscope Version No.: V 1.3 Si

Strany 2 - Declaration

2 2.13.4 Self Calibration...102 2.13.5 Self Test...

Strany 3 - Characteristic:

92 2.13 Utility System “UTILITY”: The Utility System Button. Utility Setup button Picture 2.13-1 Table 2-48 Utility System f

Strany 4

93 Picture 2.13-2 Table

Strany 5 - General Safety Summary

94 Table 2-50 Functional Menu of the Utility System 3: Option Setting Introduction Update firmware You can update the o

Strany 7

96 have a higher request to the cable , to electricity characteristic of the port. When Higher Speed connect fail, you can try to return to

Strany 8 - Preface

97 2.13.1 System status Pre

Strany 9 - Catalogue

98 Picture 2.13-7 2.13.3 Print The scope support PictBridge compatible printers. You can connect the back USB port of the oscilloscope to the US

Strany 10

99 Picture 2.13-8 Tabl

Strany 11 - Chapter 1 Accidence

100 Picture2.13-9 To print screen picture, follow these steps: 1. Connect the oscilloscope to a PictBridge compatible printer. 1) I

Strany 12 - 1.1.1 Front Panel

101 settings of your printe

Strany 14

102 2.13.4 Self Calibration Self Calibration procedure can optimize the signal path at the most measurement precision..You can run this procedure

Strany 15

103 “Single” for test.

Strany 16 - 1.2 Function checking

104 Picture 2.13-12 3. LED test ● Select “LED Test” to enter the lighten interface, the on-screen lathy rectan

Strany 17

105 2.13.6 Updating the Sys

Strany 18 - 1.3 Probe

106 9. Gate Way:10.11.0.1 10. Press‘Save/Recall’button and ‘single’button 11. name the correlated model upgrade program as ‘Upgrade.ads’,and add

Strany 19 - 1.3.3 Probe Compensation

107 Picture 2.13-15 Table

Strany 20

108 Table 2-58 Mask Setting function menu 1: Option Setting Instruction X Mask xdiv Turn the “Universal” knob to set the range of horizo

Strany 21

109 Picture 2.13-18

Strany 22 - 2.1 Menu and control button

110 option button to set the output option. 13) Enter the first page of “Pass/Fail function menu”, press the “Operate” option b

Strany 24 - 2.2 Connector

4 1.1 Accidence of Panel and Display Information 1.1.1 Front Panel It is important for you to understand the DSO’s front panel before

Strany 25 - 2.3 Auto setup

112 Picture 2.13-21 Table 2-62 Waveform play back function menu 2: Option Setting Instruction Start Frame Set start frame. Curr_Frame

Strany 26

113 menu”. 6. Select the “

Strany 27 - 2.4 Default setup

114 2.13.9 Recorder The waveform recorder is a kind of seamless and no-gap real time recording of waveform, means oscilloscope can sa

Strany 28 - 2.5 Universal knob

115 Table 2-67 recorder’s

Strany 29 - 2.6 Vertical System

116 Recorder Operation step: 1. Press the UTILITY button to enter “Utility menu”. 2. Press the “Next Page” option button to ente

Strany 30

117 2.14 Online Help Functi

Strany 31

118 Chapter3 Application Examples This section presents a series of application examples. These simple examples highlight the features of the oscill

Strany 32

119 3.1Taking Simpl

Strany 33

120 The corresponding icon and measure value will display on the third option area. ⑵⑵⑵⑵ Measure signal peak-to-peak amplitud

Strany 34

121 3.2 Taking Curs

Strany 36

122 3.2.2 Measuring Ring Amplitude To measure the amplitude, follow these steps: 1). Press the “CURSORS” button to see the Cursor Menu. 2). Press th

Strany 38 - 1. FFT Spectrum Analyzer

124 3.4 Analyze the signal detail You have a noisy signal displayed on the oscilloscope and you need to know more about it. You suspect that the sig

Strany 39

125 3.5 Triggering

Strany 40

126 3.6. Application of X-Y function Viewing Impedance Changes in a Network Connect the oscilloscope to monitor the input and output of the circ

Strany 41

127 DCA B P

Strany 42

128 3.7 Analyzing a Differential Communication Signal You are having intermittent problems with a serial data communication link, and you suspect p

Strany 43

129 Chapt

Strany 44 - 2.6.4 Using Ref

130 USB Flash Drive to the USB Host port. USB Flash Drive Pull Out! : This information will appear when you pull out the USB Flash Drive. Store

Strany 46 - 2.7 Horizontal System

6 5. Power Shutoff Switch and Input Connector Picture 1.1-4 Back and Side panel connection locations on SDS1000C/CM/CE Seri

Strany 47

132 Check the trigger mode on the trigger interface whether or not in the “normal” or “single”, and check the trigger level is whether o

Strany 49

134 5.2 Contact with us MTR Add : 3//F, Bldg No.4, Antongda Industrial Zone, 3rd Liuxian Road ,Bao’an District, Shenzhen, 518101, P.R.

Strany 50 - 2.8 Trigger System

135 Appendix A: Specifications All specification applies to 10X

Strany 51 - 2.8.2 Trigger Type

136Probe attenuator Factors Set 1X、10X、100X、1000X Vertical System Vertical Sensitivity SDS1000C/CE/CF/CFL 2mV/div - 5V/div(1-2-5 order) SDS1000CM

Strany 52

137 <3.5ns (SDS1102C/CM/CE,SDS1104CF/CFL) <5.8ns

Strany 53

138 Channel Mode Samping Rate Short memory Long Memory Single Channel 1Gsa/s 40kpts No Support Single Channel 500MSa/s or lower 20kpts 2Mpts Me

Strany 54

139 X-Y Mode X-pole Input / Y-Pole Input Channel 1 (CH1)

Strany 55

140 Display System Display Mode Color TFT 5.7in.(145mm)diagonal Liquid Crystal Display Resolution 320 horizontal by 234 vertic

Strany 56

141 Gen

Strany 58

142 Appendix B: Default setup Menu or system options、、、、 knobs or buttons Default setup Coupling DC BW limit Off Volts/div Coarse Probe 1X

Strany 59

143 Persi

Strany 60

144 Type Alternative Source CH1 Mode Edge TRIGGER (Alternative) Coupling DC

Strany 61

145 Appen

Strany 62

146 Index - Width 75 + Width 75 Accessories 122 Acquiring Signals system 57 Aliasing 60 Application of X-Y function 107 AUTO 15 auto setup 1

Strany 63

8 indicates the filter is On. 11. Readout shows the main time base setting. 12. Readout shows the horizontal positon of main timebase waveform. 13.

Strany 65 - 2.8.5 Slope & Level

10 1.3 Probe 1.3.1 Probe Safety A guard around the probe body provides a finger barrier for protection from electric shock. Picture 1.3-1

Strany 66 - Trigger level

11 Be sure that the attenuation switch on the probe matches the Pr

Strany 67 - 2.9 Acquiring Signals system

II Declaration Copyright © by Siglent Technologies Co,.Ltd. All rights reserved. Contents in this Manual are not allowed to copy, extract and tra

Strany 68

12 Over Compensated under Compensated

Strany 69

13 Chapter 2 Functions Instruction and Operation To use your o

Strany 70 - Apparent low-frequency

14 2.1 Menu and control button Showing as the following picture: Picture 2-1 All models: Channel buttons(CH1,CH2,CH3,CH4): P

Strany 71

15 SAVE/RECALL: Press to display the Save/Recall menu. You can

Strany 72 - 2.10 Display System

16 2.2 Connector Picture 2-2 Channel Connector (CH1,CH2,CH3,CH4): Input connectors for waveforms display. EXT TRIG: Input connector for an

Strany 73

17 2.3 Auto set

Strany 74

18 ● No signals found and no channels displayed, oscilloscope displays and uses channel 1. Table 2-2 Auto set the functio

Strany 75

19 2.4 Default setup The o

Strany 76 - 2.10.1 X-Y Format

20 2.5 Universal knob Picture 2-5 Universal knob You can use the Universal knob with many funct

Strany 77 - 2.11 Measure System

21 2.6 Vertical System T

Strany 78 - 2.11.2 Cursor Measurement

III SDS1000 Series Brief Introduction Model: SDS1000C Series: SDS

Strany 79

22 Table 2-4 CH1,CH2 ,CH3,CH4 function menu 2: Table 2-5 Digital Filter function menu: Option Setting Instruction Invert on off Turn on i

Strany 80 - Track mode

23 “GND” Coupling: Use GN

Strany 81

24 Setting up CH1,CH2,CH3,CH4 Channels Each channel has its own separate Menu. The items are set up separately according to each chan

Strany 82 - 2.11.3 Auto Measurement

25 frequency component whic

Strany 83

26 Picture 2.6-4 4.Setting Probe Attenuation In order to assort the attenuation coefficient, you need to response in the channel

Strany 84

27 5. Inverting waveforms T

Strany 85

28 ● Press “CH1”→ “Next Page page1/2”→ “Filter” → “On”. Turn on the Digital Filter function. Picture 2

Strany 86

29 changes the Volts/Div sc

Strany 87

30 Picture 2.6-9 1. FFT Spectrum Analyzer The FFT process mathematically converts a time-domain signal into its frequen

Strany 88 - 2.12 Storage System

31 Table 2-9 FFT function m

Strany 89

IV ● Cursor measure types: Manual mode, Track mode and Auto mode. ● Channel waveform and its FFT waveform display on split screen. ● Waveform Intensi

Strany 90

32 2. Displaying the FFT Spectrum Press the MATH button to display the Math Menu. Use the options to select the Source channel, Window algorithm, a

Strany 91 - SAVE/RECALL SETUP

33 need to use. Table 2-

Strany 92

34 5. Measuring an FFT Spectrum Using Cursors You can take two measurements on FFT spectrums: magnitude (in dB) and frequency (in Hz). Magnitude is

Strany 93

35 Picture 2.6-11 2. Mea

Strany 94

36 2.6.4 Using Ref The reference control saves waveforms to a nonvolatile waveform memory. The reference function becomes availabl

Strany 95 - SAVE/RECALL WAVEFORM

37 Press the Ref button to

Strany 96

38 2.7 Horizontal System As follow Picture, there are one button and two knobs in the HORIZONTAL area. Pictu

Strany 97

39 MemDepth Detailed Speci

Strany 98

40 display updates from left to right. There is no trigger or horizontal position control of waveforms during scan mode. 2.7.2 Window Zone Use th

Strany 99 - Save/Recall CSV

41 4. Press the “Window” b

Strany 101 -

42 2.8 Trigger System The trigger determines when the oscilloscope starts to acquire data and display a waveform. When a trigger is set up properly,

Strany 102

43 sweep speed selected. 2.

Strany 103

44 Normal Use this mode when you want to see only valid triggered waveforms; when you use this mode, the oscilloscope does not display a waveform

Strany 104

45 2. Set up Source

Strany 105 - 2.13.1 System status

46 When (Positive pulse width less than pulse width setting) (Positive pulse width larger than pulse width setting) (Positive pulse width equal

Strany 107

48 Picture 2.8-5 Table 2-18 Functional Manu of Video Trigger 2:

Strany 109

50 Table 2-19 Slope trigger function menu 1 Picture 2.8-7 Table 2-20 Slope trigger function menu 2 Option

Strany 111

VI ● Avoid Exposed Circuitry. Do not touch exposed connections and components when power is present. ● Do Not Operate With Suspected Failures. If yo

Strany 112

52 6. Press the “When” option button to select “ ”, “ ”, “ ”, “ ”, “ ”或 “ ”. 7. Press the “Time” button, turn the “Universal” knob to adjust sl

Strany 113

53 Table 2-21 Set trigger m

Strany 114 - 2.13.7 Pass/Fail

54 Table 2-23 Set trigger mode to pulse trigger function menu 2: Table 2-24 Set trigger mode to video trigger function menu1:

Strany 115

55 Table 2-26 Set trigger m

Strany 116

56 4. Press the “Type” option button to select “Alternative”. 5. Press the “Channels” option button to select “CH1-CH2” 6. Press the “Source” opt

Strany 117

57 2.8.5 Slope & Level

Strany 118 - 2.13.8 Waveform Recorder

58 Holdoff timeTrigger positionTrigger level Picture 2.8-11 If you want to change holdoff time, please follow next operations: 1. Press the “TRIG

Strany 119

59 2.9 Acquiring Signals

Strany 120

60 Sampling: In this acquisition mode, the oscilloscope samples the signal in evenly spaced intervals to construct the waveform. This mode accurat

Strany 121

61 Average: The os

Strany 123

62 the trigger point. This is also called the pre-trigger. 2). Continue to acquire data while waiting for the trigger condition to occur. 3). Detect

Strany 124

63 You can also turn Sins/x

Strany 125 - 2.14 Online Help Function

64 2.10 Display System The display function could be expressed by the “DISPLAY” Button. Display System Picture 2.10-1 Table 2-2

Strany 127

66 Picture 2.10-3 Table 2-31 Display system function menu3: Operate Introduction: 1. Set up waveform display type 1) Pr

Strany 129

68 8. Set Skin Press the “skin” option button or turn the “Universal” knob to select “Classical”, “Modern”, “Traditional” or “Succinc

Strany 130

69 ● Turn the “S/div” knob

Strany 131 - 3.3 Catch the single signal

70 maximum values of a waveform and knew you had a scale factor of 100mV/div, then you could easily calculate your peak-to-peak voltage as follows:

Strany 132

71 parallel cursors to m

Strany 133

VIII Preface This manual contains operating information for the SDS1000 Series Digital Storage Oscilloscopes. The manual consists of the following

Strany 134

72 The reciprocal of time increment between Cursor A and Cursor B: 1/△T Picture 2.11-2 Track mode Table 2-33 Track

Strany 135

73 3. Press the “Cursor A”

Strany 136

74 Picture 2.11-3 Auto mode Table 2-34 Auto mode function menu: Option Setting Instruction Cursor Mode Auto Set to au

Strany 138

76 ROVShoot, FOVShoot, RPREShoot, FPREShoot parameter. 、 、 、 、 、 、 、 、 、 、 、 、 、 Display the corresponding icon and measure value of your s

Strany 139 - 4.2 Troubleshooting

77 Off measure voltage par

Strany 140

78 + Wid + Width Measures the time between the first rising edge and the next falling edge at the waveform 50% level. - Wid -Width Measures th

Strany 141 - 5.1 Maintain summary

79 Picture 2.11-7 6 ) . P

Strany 142 - 5.2 Contact with us

80 2.12 Storage System Showing as following picture, The SAVE/RECALL is the Storage System Functional Button. Storage sys

Strany 145

82 Creating Folders and Files Creat new folders and files by pressing the New Folder or New File option button. The following screen is shown. Pict

Strany 146

83 SAVE/RECALL SETUP Sav

Strany 147

84 3). Press the “Save to” option button to select “Device”. 4). Press the “Setup” option button to select “No.1”. 5). Press the “DISPLAY” button to

Strany 148

85 Picture 2.12-7 Save se

Strany 149

86 Recall setup data from USB flash drive, follow next steps: 1) Press the “SAVE/RECALL” button. 2) Press the “Type” button to select “Setups”. 3) I

Strany 151

88 want to need. 7). Press the “Save” option button. Recall wavefroms, follow next steps: 1). Press the “SAVE/RECALL” button to enter “SAVE/RECALL”

Strany 152

89 the oscilloscope

Strany 153 - Cleaning

90 Picture 2.12-11 Save waveform image to USB flash drive, follow next steps: 1). Select the screen image that you want . 2) Press the “SAVE/RECALL

Strany 154

91 Save/Recall CSV