Service Manual
Specification Test
Purpose: Test if some specifications are in the specified range.
Table 3-2 require instruments
Description Minimum requirements Examples
DC Voltage Source
-6 mV to 30 V, ±0.1% accuracy
Standard sine wave
generator
10Hz and 300MHz, ±1% amplitude
accuracy
FLUKE 9500B
50 Ω BNC
Feed-Through
Termination
DC to 1GHz PL-50
Before beginning these procedures, two conditions must first be met:
● The oscilloscope must have been operating continuously for thirty minutes within the
operating temperature range specified.
● You must do the Self Calibration procedure. If the ambient temperature changes by more
than 5℃, you must do the Self Calibration procedure again.
Picture 3-9 Test instrument connecting
Test DC Gain Accuracy
Steps: 1、Set the DC voltage source output level to 30 V
2、Press the button “DEFAULT/SETUP” of oscilloscope
3、Set VOLT/DIV to 10V/div, set measure parameter as “CH1,Mean”
4、Connect the oscilloscope channel 1 to the DC voltage source, and record the value of
Mean appeared on the screen as V
pos
.
5、Change the polarity of the DC voltage source,and record the value of Mean appeared
on the screen as V
neg
.
6、Calculate V
diff
=V
pos
– V
neg
,then compare V
diff
to accuracy limits in the table.